In this study, we performed a numerical analysis of a novel EMF Birdcage wearable device for the treatment of Alzheimer’s disease (AD). We designed the new device to generate and radiate a frequency of 64 MHz and a specific absorption rate (SAR) of 0.6 W/kg to a simulated human brain tissue. We determined these parameters from our experimental studies on primary human brain cultures at the Indiana University School of Medicine (IUSM). We found that this frequency and SAR decreased the toxic Aβ levels in the cell cultures. The frequency of 64 MHZ has good skin depth penetration, which will easily pass through the various head layers, including hair, skin, fat, dura, the cerebrospinal (CSF), and grey matter, and reach deeply into the brain tissues. The SAR of 0.6 W/kg was achieved with lower power input and energy, decreasing the probability of thermal injury. Therefore, these parameters enhance the safety of these potential treatments. This Birdcage device emulates a small-scale MRI machine, producing the same 64 MHz frequency at much lower operating input power. In this work, we utilized a high-frequency simulation system (HFSS/EMPro) software to produce the birdcage structure for the required EMF parameters. The 64 MHz radiating frequency produced the scattering S11 parameter of −15 dbs. We obtained a SAR of 0.6 W/kg when an input power of 100 W was applied. The coil dimensions were found to be near 15 cm in height and 22 cm in diameter, which fits in wearable systems. We found that the distribution of the electric field and SAR radiate homogeneously over the simulated human head with good penetration into the brain, which proves to be an appropriate po-tential therapeutic strategy for Alzheimer’s disease.